LabDELTA™
Solves Laboratory Waste
Treatment Problems |
|
|
|
 |
 |
|
LabDELTA™ from Wastech
automates laboratory liquid processes
from
waste treatment to dispensing and chemical mixing.
LabDELTA™ from Wastech Controls solves laboratory waste treatment problems
with small-volume, automated systems. These are ideal as
school laboratory
waste treatment systems. Complete packages are available in
either 2, 4 or 6 gallon batch designs or 1, 2 or
3 gpm
continuous designs with complete
controls, including communication options for alarm reporting. Acid waste
neutralization (AWN) including
strong acid neutralization (0-2 pH) is typical. Other
systems handle heavy metals, fluoride reduction, and pH adjustment for
caustic fluids, automatically and without risk to lab personnel. Accidents from
exothermic reactions, spills, burns and glass breakage are eliminated.
Using various system designs including two-stage neutralization,
limestone pretreatment, or packaged chillers with PVDF or Telfon heat
exchangers, Wastech Controls treats the waste safely and prepares it for
dumping into the sewer. The units are small and compact, fabricated of
polypropylene, with drip pans or double-containment as requested, and
include all sensors and controllers.
|
 |
 |
|
Clean control
cabinet wiring. |
PP heat exchanger at bottom of reaction
tank. |
Wastech Controls has been supplying waste treatment
solutions since 1987, and brings this design
experience into the lab with reliable designs and durable components such as
pH sensors and pH controllers.

Wastech's
LabDELTA™ laboratory waste treatment units are
quiet, sealed to eliminate vapors, and designed to be
vented to building systems. Solvent waste, acid waste,
and most lab liquid wastes are handled automatically. Pump options
incorporating high level alarms and automatic shut-offs
are available to transfer untreatable waste to drums.
The units can be monitored locally with alarm
lights, or remotely on a PC using web-based TCP/IP
communications or modem-to-modem dial-up.
LabDELTA units are small enough to be located in the laboratory itself.
They can include collection tanks with pumps or be
located near floor drains for gravity discharge into the
sewer. The units can be located under counters or in
cabinets. Control panels are normally located on the
units, but can be remotely mounted anywhere in the lab.

LabDELTA Extreme for
Laboratory Waste Treatment for Strong Acids & Caustics
Packaged Laboratory Waste Treatment Systems
Batch Systems
Part
Number |
Batch Size
(gallons) |
Neutralization Tank (gallons) |
Chemical
Tank (gallons) |
Mixer
(HP) |
Metering
Pump (GPH) |
|
LD02 |
2 |
2 |
2 |
1/20 |
0.6 |
|
LD04 |
4 |
6 |
2 |
1/20 |
1.0 |
|
LD06 |
6 |
11 |
2 |
1/20 |
1.0 |
|
LDE006* |
6 |
2 |
2 |
1/20 |
0.6 |
Continuous Systems
Part
Number |
Flow Rate
(GPM) |
Neutralization Tank (gal) |
Monitoring Tank (gal) |
Chemical
Tank (gal) |
Mixer
(HP) |
Metering
Pump (GPH) |
|
LD10 |
1 |
33 |
22 |
11 |
1/20 |
1.0 |
|
LD20 |
2 |
42 |
28 |
15 |
1/20 |
1.0 |
|
LD30 |
3 |
50 |
33 |
36 |
1/20 |
1.0 |
|
|